1560 Black Stack Thermometer Readout
1560 Black Stack Thermometer Readout is one of the most versatile, cost-effective & accurate readouts that you are looking for.
The 1560 Black Stack thermometer readout can be any kind of thermometer you want it to be, and it works in three distinctive ways.
It’s a reference thermometer with an accredited calibration; it’s an automated calibration system reading your reference probe and sensors you’re testing; or it’s a high-accuracy data acquisition system. And it does each of these functions better than any other thermometer currently on the market.
The Stack consists of up to eight different modules that fit together to do any type of thermometry you choose. You can buy all of them, or any combination of them, and change the Stack and its functions anytime you want. Each module stacks behind the preceding one, and when you add a module, the Stack’s software automatically reconfigures itself to include all of the new functions supplied by that module. There’s nothing to take apart. No boards need to be installed.
There’s no software to load, and nothing has to be calibrated. Just stack a new module onto the back of the previous modules and you’re ready to use the Black Stack and all of its remarkable features.
Key features: 1560 Black Stack Thermometer Readout
- Reads SPRTs, RTDs, thermistors, and thermocouples
- Any configuration you like up to eight modules
- High-accuracy reference thermometer (to ±0.0013°C)
- Automates precision data acquisition
Specifications: 1560 Black Stack Thermometer Readout
Specifications |
|
Model 1560 Base Unit | Power: 100 to 240 VAC, 50 or 60 Hz, nominal; Attachable Modules: up to 8; Display: 4.25" x 2.25" LCD graphics, LED backlight, adjustable contrast and brightness; Automatic Input Sequencing: 1 to 96 channels; Communications: RS-232; Non-volatile Memory: channel sequence, probe coefficients; Minimum Sample Time: 2 seconds. |
Extended Communication Module 3560 | The Extended Communication Module adds additional communication interface capability to the system. This module includes a GPIB (IEEE-488) interface, Centronics printer interface, and analog output. The GPIB interface connects the 1560 to a GPIB bus. GPIB can be used to control any function of the 1560 and read measurement data. The printer interface allows the 1560 to send measurement data directly to a printer. The analog output sources a DC signal (±1.25 VDC) corresponding to the value of a measurement. |
Resistance Modules | ||||||||
. |
Input Channels
|
Resistance Range
|
Basic Resistance Accuracy
|
Resistance Resolution
|
Temp.
Range |
Equivalent Temperature Accuracy†
|
Temp.
Resolution |
Excitation Current
|
SPRT Module 2560 |
2
|
0Ω to 400Ω
|
±20 ppm of reading (0.0005Ω at 25Ω, 0.002Ω at 100Ω)
|
0.0001Ω
|
–260°C to 962°C
|
±0.005°C at 0°C ±0.007°C at 100°C
|
0.0001°C
|
1.0 mA,
1.4 mA |
High-Temp PRT Module 2561 |
2
|
0Ω to 25Ω
|
±50 ppm of reading (0.00013Ω at 2.5Ω)
|
0.00001Ω
|
0°C to 1200°C
|
±0.013°C at 0°C ±0.018°C at 100°C
|
0.001°C
|
3.0 mA,
5.0 mA |
PRT Scanner 2562 |
8
|
0Ω to 400Ω
|
±40 ppm of reading (0.004Ω at 100Ω)
|
0.0001Ω
|
–200°C to 850°C
|
±0.01°C at 0°C ±0.014°C at 100°C
|
0.0001°C
|
1.0 mA,
1.4 mA |
Standards Thermistor Module 2563 |
2
|
0Ω to 1 MΩ
|
±50 ppm of reading (0.5Ω at 10 KΩ)
|
0.1Ω
|
–60°C to 260°C
|
±0.0013°C at 0°C ±0.0015°C at 75°C
|
0.0001°C
|
2 µA,
10 µA |
Thermistor Scanner 2564 |
8
|
0Ω to 1 MΩ
|
±100 ppm of reading (1Ω at 10 KΩ)
|
0.1Ω
|
–60°C to 260°C
|
±0.0025°C at 0°C ±0.003°C at 75°C
|
0.0001°C
|
2 µA,
10 µA |
1000Ω PRT Module 2567 |
2
|
0Ω to 4 KΩ
|
±25 ppm of reading (0.025 at 1 KΩ)
|
0.001Ω
|
–260°C to 962°C
|
±0.006°C at 0°C ±0.009°C at 100°C
|
0.0001°C
|
0.1 mA,
0.05 mA |
1000Ω PRT Scanner 2568 |
8
|
0Ω to 4 KΩ
|
±40 ppm of reading (0.04 at 1 KΩ)
|
0.001Ω
|
–200°C to 850°C
|
±0.01°C at 0°C ±0.014°C at 100°C
|
0.0001°C
|
0.1 mA,
0.05 mA |
Thermocouple Modules | |||||||
. |
Input Channels
|
Millivolt Range
|
Millivolt Accuracy
|
Millivolt Resolution
|
Temperature Accuracy,† Ext. CJC
|
Temperature Accuracy,† Int. CJC
|
Temperature Resolution
|
Precision Thermocouple Module 2565 |
2
|
–10 to 100 mV
|
±0.002 mV
|
0.0001 mV
|
±0.05°C
|
±0.1°C
|
0.001°C††
|
Thermocouple Scanner 2566 |
12
|
–10 to 100 mV
|
±0.004 mV
|
0.0001 mV
|
±0.1°C
|
±0.3°C
|
0.01°C
|
†Temperature accuracy depends on probe type and temperature ††Actual temperature resolution may vary depending on measurement noise, thermocouple sensitivity, and input average |
Includes: 1560 Black Stack Thermometer Readout
Model Name | Description |
---|---|
1560 |
Black Stack Readout Base Unit |
2560 |
SPRT Module, 25 Ω and 100 Ω, 2-channel |
2561 |
High-Temp PRT Module, 0.25 Ω to 5Ω, 2-channel |
2562 |
PRT Scanner Module, 8-channel |
2563 |
Standards Thermistor Module, 2-channel |
2564 |
Thermistor Scanner Module, 8-channel |
2565 |
Precision Thermocouple Module, 2-channel |
2566 |
Thermocouple Scanner Module, 12-channel |
2567 |
SPRT Module, 1000 Ω, 2-channel |
2568 |
PRT Scanner Module, 8-channel, 1000 Ω |
3560 |
Extended Communications Module |
5610-6-X |
Thermistor Probe, 3.2 mm dia x 152.4 mm (0.125 in x 6 in), 0 °C to 100 °C, Lead wire length (Cable) 6 ft (Calibration traceable to NIST standards included) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5610-9-X |
Thermistor Probe, 3.2 mm dia x 228.6 mm (0.125 in x 9 in), 0 °C to 100 °C, Lead wire length (Cable) 6 ft (Calibration traceable to NIST standards included) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5626-12-X |
Secondary Standard PRT (0.25 in dia x 12 in), 100 Ω, –200 °C to 661 °C (17025 accredited calibration included. Traceable to NIST standards.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5626-15-X |
Secondary Standard PRT (0.25 in dia x 15 in), 100 Ω, –200 °C to 661 °C (17025 accredited calibration included. Traceable to NIST standards.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5628-12-X |
Secondary Standard PRT (0.25 in dia x 12 in), 25 Ω, –200 °C to 661 °C (17025 accredited calibration included. Traceable to NIST standards.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5628-15-X |
Secondary Standard PRT (0.25 in dia x 15 in), 25 Ω, –200 °C to 661 °C (17025 accredited calibration included. Traceable to NIST standards.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5642-X |
Standards Thermistor Probe (Calibration traceable to NIST standards included) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
Accessories common to all models:
Accessory | Description |
---|---|
9302 |
Rugged Carrying Case for these models:
|
2380-X |
Miniature Thermocouple Connectors |
2381-X |
Standard Thermocouple Connector |
2382 |
RTD/Thermistor Connector |
1929-2 |
System Verification, PRT with Readout, Accredited. |
1929-2 |
System Verification, PRT with Readout, Accredited. |
1929-5 |
System Verification, Thermistor with Readout, Accredited. |
1930 |
System Calibration, PRT with Readout, Accredited. |
1935 |
System Calibration, Thermistor with Readout, NVLAP-accredited |
2560 |
SPRT Module, 25 and 100 2-channel |
2562 |
PRT Scanner Module, 8-channel |
2566 |
Thermocouple Scanner Module, 12-channel |
2568 |
PRT Scanner Module, 8-channel, 1000 Ω |
3560 |
Extended Communications Module |
9935-S |
Software, LogWare II, Single User |
9938 |
MET/TEMP II Software |
Priority Gold Instrument CarePlan |
Fluke Calibration Priority Gold Instrument CarePlans are available for most calibration products. |
Silver CarePlan |
Fluke Calibration Silver CarePlans are available for most calibration products. Please contact your local Fluke Calibration sales representative for details or to request a quote. |